BeiJing Cape Golden Gas System Company LTD yelling@golden-cape.com
CI-PC96 Process Trace Oxygen Analyser Trace O2 Analyzer 100ppm~21%

CI-PC96 Process Trace Oxygen Analyser Trace O2 Analyzer 100ppm~21%

  • High Light

    CI-PC96 trace oxygen analyser

    ,

    trace oxygen analyser 100ppm 21%

    ,

    trace o2 analyzer 21%

  • Sample Gas Flow
    150~300ml/mim
  • Accuracy
    ±2% F.S
  • Ambient Temp
    -10℃~+50℃
  • Stability
    <±1%.FS (168h)
  • Ambient Humidity
    <80%RH
  • Signal Output
    4~20mA
  • Measure Range
    10ppm~21%, 100ppm~21%, 1000ppm~21% O2
  • Range
    10.00~96.00% O2·
  • Place of Origin
    BeiJing
  • Brand Name
    cape golden
  • Certification
    CE ISO
  • Model Number
    CI-PC96
  • Minimum Order Quantity
    1
  • Delivery Time
    15 day
  • Payment Terms
    L/C, D/A, D/P, T/T, Western Union, MoneyGram

CI-PC96 Process Trace Oxygen Analyser Trace O2 Analyzer 100ppm~21%

CI-PC96 Process Trace Oxygen Analyzer

 

CI-PC96 trace O2 analyzer is with electrochemical principle and capable of O2 purity measurement range 0-10/100/1000ppm, 1%.

Features

  • 240×128 LCD display oxygen concentration and other settings;
  • Isolated 4-20mA signal output, with 0-100mV, 0-1V and 0-10V as options;
  • Easy to replace sensor;
  • Standard RS232 two-way communication port;
  • Two way SPST alarms;
  • Over pressure protection function and over purity range protection function;
  • Automatic data storage in a certain interval time, and browsing data at any time

Technical Specifications

Model: CI-PC96

Measure Range: 0-10/100/1000ppm, 1% O2

Principle: Electrochemical

Measure Mode: Online

Install: Dial mounted

Application: Non dangerous general applications

Sensitivity: 0.5% F.S.

Measure Accuracy: ≤±5% @ 0-10ppm ; ≤±3% @ 0-100ppm; ≤±2% @0-1000ppm

Stability: ≤±2.5% @ 0-10ppm; ≤±1.5% @ 0-100ppm; ≤±1% @ 0-1000ppm

Response Time: T90<60s under 25℃

Sample Gas Temperature: 0-45℃

Sample Gas Flow: 1.2L/min

Sample Gas Pressure: 35-210KPa

Power Supply: AC85~265, consumption lower than 20VA

Signal Output: 4-20mA DC isolated, with 0-100mV, 0-1V and 0-10V to switch

Alarms: 2 way SPST alarm

Alarm Contact Capacity: 0.2A under AC240V

Display Resolution: 0.01%@%; 1ppm@100-10000ppm; 0.1ppm@10-100ppm; 0.01ppm@0-10ppm

Communication: RS232 standard (RS485 optional)

Analyzer Size: 144 (W)×144 (H)×190 (D) mm

Applications

Air separation, Gases production, Protection gas for semi-conductors industry, Wave soldering & reflow soldering, Food production, Petro-chemical, and other oxygen purity detection in inert and hydrocarbon gas mixtures.